SENSORS, METROLOGY GRADE, 100 ohm
PLATINUM RTD
(for Instrulab
Models 4200C and 4312A or any instrument having ITS-90 capability)

Laboratory Grade, 830 series
Temperature
range: -100° to +400°C (-150° to +750°F)
Transition fitting maximum temperature: 260°C
Selection
Guide |
Model |
Diameter |
Length |
A |
B |
C |
Response time
(seconds) |
833 |
1/8" |
8" |
1" |
1 1/2" |
3/8" |
3 |
831 |
3/16" |
10" |
1" |
1 1/2" |
3/8" |
5 |
832 |
1/4" |
10" |
1" |
1 1/2" |
3/8" |
7 |
Accuracy
(with custom calibration) |
Celsius |
-100°C |
0°C |
100°C |
200°C |
300°C |
400°C |
|
±0.036 |
±0.012 |
±0.017 |
±0.023 |
±0.035 |
±0.150 |
Fahrenheit |
-150°F |
32°F |
212°F |
390°F |
575°F |
750°F |
|
±0.068 |
±0.023 |
±0.032 |
±0.044 |
±0.067 |
±0.285 |
Secondary Standard, Model 840
Temperature
range: -200° to +500°C (-328° to +932°F)
Transition fitting maximum temperature: 150°C
Mechanical
Specifications |
Model |
Diameter |
Length |
A |
B |
C |
Response time
(seconds) |
840 |
1/4" |
12" |
3 3/4" |
2 3/4" |
1/2" |
9 |
Calibration
Uncertainty in degrees (with custom calibration) |
Celsius |
-200°C |
-100°C |
0°C |
100°C |
200°C |
300°C |
400°C |
500°C |
|
±0.028 |
±0.036 |
±0.012 |
±0.017 |
±0.023 |
±0.029 |
±0.037 |
±0.100 |
Fahrenheit |
-328°F |
-150°F |
32°F |
212°F |
390°F |
575°F |
750°F |
932°F |
|
±0.053 |
±0.068 |
±0.023 |
±0.032 |
±0.044 |
±0.055 |
±0.070 |
±0.190 |
Documentation supplied with custom calibrated sensors includes:
- Date, model serial number of the sensor.
- Measured temperature and reading taken, in ohms.
- Calibration constants; ITS-90 coefficients. All required coefficients required to
program the Instrulab 4200 Series instruments and the Model 4312A instrument.
- Resistance vs. Temperature table in printout increments of 5° or 10° as required.
Resistance resolution of 0.001 ohms.
|