SYSTEMS UNCERTAINTY +
°C
WORST CASE (INSTRUMENT + SENSOR)
|
INSTRUMENT MODEL |
SENSOR MODEL & RANGE NBR |
NBR CHANNEL(S) |
RESOLUTION ° C or °F |
RANGE
NOTE 1
|
-200°C |
-100°C |
0°C |
+100°C |
200°C |
300°C |
400°C |
500°C |
660°C |
962°C |
|
-328°F |
-150°F |
32°F |
212°F |
390°F |
575°F |
750°F |
932°F |
1220°F |
1765°F |
|
4201C |
832-153-01 |
1 |
0.01° |
-100 to +400°C |
- |
±0.05°C |
±0.02°C |
±0.03°C |
±0.04°C |
±0.05°C |
±0.21°C |
- |
- |
- |
|
4202C |
" |
2 |
" |
" |
- |
" |
" |
" |
" |
" |
" |
- |
- |
- |
|
4212C |
" |
2 + Delta T |
" |
" |
- |
" |
" |
" |
" |
" |
" |
- |
- |
- |
|
4201C-27 |
832-153-01 |
1 |
0.01° |
-100 to +400°C |
- |
±0.045°C |
±0.015°C |
±0.026°C |
±0.037°C |
±0.049°C |
±0.211°C |
- |
- |
- |
|
4201C-27 |
" |
2 |
" |
" |
- |
" |
" |
" |
" |
" |
" |
- |
- |
- |
|
4201C-27 |
" |
2 + Delta T |
" |
" |
- |
" |
" |
" |
" |
" |
" |
- |
- |
- |
|
4312A |
" |
|
" |
" |
- |
±0.047°C |
±0.018°C |
±0.028°C |
±0.039°C |
±0.052°C |
±0.214°C |
- |
- |
- |
|
4201C-33 |
832-153-01 |
1 |
0.001° |
-100 to +400°C |
- |
±0.043°C |
±0.014°C |
±0.024°C |
±0.035°C |
±0.046°C |
±0.206°C |
- |
- |
- |
|
4202C-33 |
" |
2 |
" |
" |
- |
" |
" |
" |
" |
" |
" |
- |
- |
- |
|
4212C-33 |
" |
2 + Delta T |
" |
" |
- |
" |
" |
" |
" |
" |
" |
- |
- |
- |
|
4201C |
840-164-01 |
1 |
0.01° |
-200 to +500°C |
±0.02°C |
±0.02°C |
±0.02°C |
±0.02°C |
±0.02°C |
±0.03°C |
±0.03°C |
±0.03°C |
- |
- |
|
4202C |
" |
2 |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
- |
- |
|
4212C |
" |
2 + Delta T |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
- |
- |
|
4201C-27 |
840-164-01 |
1 |
0.001° |
-200 to +500°C |
±0.011°C |
±0.012°C |
±0.013°C |
±0.015°C |
±0.017°C |
±0.020°C |
±0.023°C |
±0.028°C |
- |
- |
|
4202C-27 |
" |
2 |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
- |
- |
|
4212C-27 |
" |
2 + Delta T |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
- |
- |
|
4312A |
" |
12 |
" |
" |
±0.013°C |
±0.014°C |
±0.016°C |
±0.017°C |
±0.019°C |
±0.023°C |
±0.026°C |
±0.032°C |
- |
- |
|
4201C-33 |
840-164-01 |
1 |
0.001° |
-200 to +500°C |
±0.009°C |
±0.010°C |
±0.012°C |
±0.013°C |
±0.015°C |
±0.017°C |
±0.018°C |
±0.022°C |
- |
- |
|
4202C-33 |
" |
2 |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
- |
- |
|
4212C-33 |
" |
2 + Delta T |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
- |
- |
|
4201C |
*162W |
1 |
0.01° |
0 to +962°C |
- |
- |
±0.33°C |
±0.33°C |
±0.33°C |
±0.33°C |
±0.34°C |
±0.34°C |
±0.34°C |
±0.36°C |
|
4202C |
" |
2 |
" |
" |
- |
- |
" |
" |
" |
" |
" |
" |
" |
" |
|
4212C |
2+Delta T |
" |
" |
" |
- |
- |
" |
" |
" |
" |
" |
" |
" |
" |
|
4221C |
**SPRT |
1 |
0.001° |
-200 to +600°C |
±0.004°C |
±0.004°C |
±0.005°C |
±0.005°C |
±0.006°C |
±0.006°C |
±0.007°C |
±0.007°C |
±0.008°C |
- |
|
4222C |
" |
2 |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
" |
- |
|
410C |
800 |
1 |
" |
0.001° |
-10° to 400 |
0.048 |
0.056 |
0.029 |
0.037 |
0.043 |
0.056 |
0.098 |
" |
" |
* 162W is a Rosemount sensor, Ro=10 ohms, calibrated uncertainty is ±0.25ºC
**Rosemount or YSI SPRT, Ro=25.5 ohms, assumes calibrated uncertainty of ±0.001ºC
Note 1 Not all available ranges have been listed. Narrower ranges, however,can still
use the uncertainties listed for any particular instrument/sensor combination.
|